Performance Evaluation of Image Analysis Methods 381
TABLE 12.4
Approximated Probabilities of Over- and Under-Detection o f the
Number of Image Regions for Images in Figure s 12.1a–l.
µ σ ∆ P
ov
(+1)
µ σ ∆ P
ud
(−1)
a −205 18.4 −473 Φ(−14.5) −906 5.57 −1752 Φ (−152)
b −94 5.49 −251 Φ(−28.6) −34 4 2.49 −627 Φ(−114)
c −68 3.26 −198 Φ(−3 9.9) −216 1.75 −371 Φ(−88.6 )
d −53 1.90 −168 Φ(−6 0.5) −133 1.21 −203 Φ(−57.9)
e −48 1.35 −159 Φ(−82.2) −101 0.94 −139 Φ(−40.4)
f −50 1.02 −163 Φ (−111.) −83. 0.78 −104 Φ(−26.9)
g −50 0.82 −163 Φ(−138.) −71. 0.68 −80. Φ(−13.2)
h −48 0.69 −159 Φ(−161.) −61. 0.61 −61. Φ(+0.00)
i −46 0.60 −155 Φ(−182.) −53. 0.55 −44. Φ(+16.4)
j −43 0.54 −148 Φ(−194.) −45. 0.51 −28. Φ(+33.3)
k −37 0.48 −137 Φ(−208.) −37. 0.49 −13. Φ(+49.0)
l −32 0.45 −126 Φ(−209.) −30. 0.46 3.0 Φ(+71.7)
a b c
d e f
FIGURE 12.4
X-ray CT image of a physical phantom (a) and its five co mponents (b ∼ f).
2) Results from a real image
Figure 12.4a shows a real X-r ay CT phantom image that is shown in Section
10.5.1. For convenience of description, some c ontents of Section 10.5.1 are
used in this section. This physical phantom consists of six cylinders (made of
four types of materials: Poly, 0 13A, Teflon, and Bone) with nearly the same
diameters. These cylinders are mounted on a base in such a way that their axes
are parallel to each other and equia ngularly located on a circle. The resolution
of the X-ray CT scanner that was used to gene rate this image is 1.5×1.5×5.0
mm
3
. While imaging, the axes o f the cylinders are set paralle l to the moving
direction of the scanner bed. Thus, the image shown in Figure 12.4.a includes
Get Statistics of Medical Imaging now with the O’Reilly learning platform.
O’Reilly members experience books, live events, courses curated by job role, and more from O’Reilly and nearly 200 top publishers.