M13.5. Multiple SISO Controllers

Setpoint Changes

In Section M13.3, the composition control loops were tuned independently using the IMC-based PI technique, neglecting time delays. The control system was simulated assuming a 2-minute time delay on all measurements. The IMC filter factors (λ1, λ2) were both set to 25 minutes, one third of the process time constant. Here, we show the responses for both loops closed simultaneously. The response for a setpoint change of 1 mol% (0.01 mole fraction) in the distillate composition is shown in Figure M13-6. Note that the controllers are highly interacting and the response time is extremely slow (the outputs are still far from the setpoint after 1000 minutes).

Figure M13-6. Setpoint change of r = [1,0], ...

Get Process Control: Modeling, Design, and Simulation now with the O’Reilly learning platform.

O’Reilly members experience books, live events, courses curated by job role, and more from O’Reilly and nearly 200 top publishers.